跳到正文
ZH

PART SALES · Thermo Fisher · PRIMA PRO - Components

MICRO CAPILLARY (HS-AAC-MC)

BrandThermo Fisher Scientific
分类PRIMA PRO - Components
Part Number204-764-006
Item NameMICRO CAPILLARY ASSEMBLY
Description
  • 该部件安装于 Sample Transfer Assembly 内。
  • 流经 Capillary 的流量进入 Inlet Probe。
  • 更换时注意不要损坏前端的 Capillary 管。

- The Micro Capillary is glued into a m5 grub screw which is located in the sample transfer assembly.

- The flow through the capillary is 7 – 10 ml/min, while the flow rate through the leak element is around 5x10-4 mbar l/s, i.e. only a small fraction (<1%) of the gas entering the system through the capillary passes to the ion source, the remainder being exhausted via the rotary pump.

- The flow through the capillary, transfer line, inlet probe and bypass line to the pump can be considered as a fast loop, so that changes in sample gas composition are rapidly transmitted to the inlet probe.